Release notes

What changed in each recent release of Allego and Videre, newest first.

Download the latest

Allego 4.1.9

Released 2026-11-19 · Release on GitHub

Features

  • Added Computer Audio: Allego now plays audio on this computer straight from the recording stream, so channel height and amplitude range no longer change the sound. A new Computer Audio dialog has band filter, noise slicer and volume controls.
  • Added an option to run the Monitor's visualization (stage-2) filters on computer audio
  • Computer audio settings are now saved separately for each Monitor tab
  • Added a noise slicer and a reference channel for the analog audio outputs (AO 1/2). Clearing an AO's site now turns that output off.
  • Added a Test Frequency setting to the Impedance tab. It is fixed at 1000 Hz on Smartbox Classic, Open Ephys USB2 and Intan USB2 instruments.
  • Made the Monitor much faster at high channel counts. At 256 channels:
    • Streaming holds a steady 60 fps, up from 10–13.
    • Scrolling with every channel visible is smooth, at 60 fps instead of 2–9.
    • Redrawing after a scroll or zoom takes about 70 ms instead of up to 1.7 s.
  • Added mouse wheel zoom in the Monitor:
    • Alt + scroll changes channel height.
    • Shift + scroll changes amplitude range, or AIN range when the pointer is over an AIN channel.
    • Ctrl + scroll or a trackpad pinch changes both.
  • The channel under the mouse pointer stays in place while you zoom channel height with the wheel
  • Added more channel height options
  • Monitor channels now keep their trace history when you hide, add or reorder channels, or toggle GPIO-on-top
  • Added tooltips that show the wheel zoom shortcuts for channel height and range
  • Reworked the Help window: a cleaner index that lists only the current app's tabs, and links that always open radiens.ai
  • Removed the check that blocked signing in from more than one device
  • Organization admins now see their own seat, the seat list refreshes after an invite, and invited users can finish setting their password inside the app
  • Pro plans include 10 seats per organization
  • Reduced memory use when reading long recordings

Fixes

  • The whole Monitor view now redraws right away when channel height, range or clipping changes, including while stopped
  • Limited the analog output gain picker to the values the hardware supports (2^0–2^7)
  • Fixed hardware filter, analog output and digital output settings being reset after a headstage scan
  • Fixed the hardware bandpass filter being lost after a sample rate change
  • Fixed headstage sensor and channel selections being lost after a scan failed right after streaming stopped. This happened mainly on Windows with XDAQ Gen2.
  • Fixed analog outputs 3–8 interfering with left audio channel routing
  • Fixed analog output high-pass settings not being reapplied after a sample rate change
  • Fixed thin dark breaks appearing in thick digital channel traces in the Monitor
  • Fixed the Monitor briefly drawing at the old scroll position while zooming channel height
  • Fixed sorts, KPI files and file types not being found for recordings with a dot in their name, such as rec.v2
  • Fixed Open Ephys recording folders with a dot in their name being opened as the wrong folder
  • Fixed MEArec imports with a dot in the file name reading or writing the wrong files
  • Fixed Monitor slowdowns caused by a background worker leaking each time the Monitor opened

Allego 4.1.5

Released 2026-11-06 · Release on GitHub

Features

  • Added a "Resend code" link to the sign-up verification screen
  • The verification and password reset screens now tell you to check your spam or junk folder for an email from noreply@radiens.neuronexus.com
  • Signing in with an account that hasn't been verified now takes you to the verification screen, with a prompt to enter or resend your code
  • The links on the sign-in and sign-up screens can now be reached and used from the keyboard

Fixes

  • Replaced raw error text on the sign-in and sign-up screens with clear messages, such as "Too many attempts. Wait a few minutes and try again." and "We couldn't send a code to that email. Check the address and try again."
  • Fixed an incorrect verification code clearing your account details and blocking another attempt

Allego 4.1.4

Released 2026-11-05 · Release on GitHub

Features

  • Added a free 14-day Pro trial from the Upgrade dialog. "Buy now" converts an active trial, and your profile shows how many days are left
  • The privacy settings and dialog now say that anonymous usage telemetry includes your license plan names

Fixes

  • Fixed Allego starting in the default mode instead of restoring your last workspace when that workspace used a licensed DAQ mode
  • Fixed new accounts sometimes showing "Expired" and being refused at startup because license setup hadn't finished yet
  • Fixed Pro features staying unlocked after a trial or subscription ended. Licenses are now rechecked at trial end and at least every hour
  • Fixed license and account details from a previous user sometimes carrying over after signing out, switching users, or a failed sign-out
  • Fixed a spike detected twice when it fell on a processing-chunk boundary
  • Fixed spikes from neighboring channels being drawn on the wrong row of the spike raster
  • Fixed a crash when the spike panel was closed while it was still loading

Allego 4.1.2

Released 2026-10-25 · Release on GitHub

Fixes

  • Fixed the Raster tab crashing when opened

Allego 4.1.1

Released 2026-10-23 · Release on GitHub

Fixes

  • Fixed AOut Controls > Zoom landing on the bottom channels of the Monitor instead of the analog out channel
  • Workspaces saved while zoomed in on an analog out channel now show their channel height and amplitude range in the Monitor dropdowns instead of leaving them blank
  • Enlarged the µV scale labels on tall Monitor channels
  • Fixed the datasources table filter panel clipping its labels and showing the value field without a border

Allego 4.1.0

Released 2026-10-22 · Release on GitHub

Features

  • Re-enabled the analog out (DAC) highpass filter on Gen2 xSR systems
  • Added the X3-H32 connector mapping for the 30-site mouse EEG probe
  • Probe position controls are now available in the Standard tier
  • Improved stability with XDAQ hardware on macOS
  • Reduced the application's install size

Fixes

  • Fixed the acquisition board becoming unreachable on macOS after a few minutes of auto-scan, until Allego was restarted
  • Fixed a crash on macOS when an XDAQ Gen2 is unplugged
  • A board that loses power and comes back now reconnects without having to toggle the system mode
  • Fixed Allego freezing when the board's USB connection drops
  • Fixed a delay when starting a stream right after a restart
  • Recordings from XsR and Smartbox systems now save the correct start time instead of January 1, 1970
  • Fixed the spike sorter silently stopping after a cable delay change or a failed workspace load; it previously still showed as on while no spikes appeared
  • Workspaces whose saved recording folder no longer exists now load, keeping the current recording folder
  • Fixed spike files and sidecars saved by earlier versions opening with a sample rate of 0 and zeroed time ranges
  • Fixed a crash that could blank the Spike Sorter tab at startup
  • Fixed the Spike Sorter showing zeros after a workspace restore while the picker showed the loaded sort
  • Fixed Sort Details listing the neurons-per-site mode and count as quartiles (Q1/Q2)
  • Fixed a PSTH error when a dataset is reloaded
  • Fixed the Monitor being blank when opened during a stream, or missing its labels, scale bars and grid when opened while stopped
  • Fixed a reopened Monitor showing the previous Monitor's time labels
  • Fixed overlapping time labels on the Monitor and the spectrogram in narrow panes
  • Fixed neuron grid titles and x-axis labels overlapping in narrow Neurons panes
  • Spectral sidebars now collapse automatically when the plot area is too narrow, and the "no data" message stays inside the plot area
  • Fixed the Data Sources summary and the Spike Sorter dashboard clipping in narrow panes
  • Fixed labels sitting indented from their controls, including "Time Range" overlapping its clock icon in the Neurons toolbar
  • Fixed Windows compatibility on systems where the WMIC tool has been removed

Allego 4.0.13

Released 2026-10-12 · Release on GitHub

Features

  • Added a ±5 mV option to the Monitor amplitude range
  • DSP filter frequency sliders now use a logarithmic scale with a tick at each decade and Hz/kHz labels, so low frequencies are easier to set
  • Redesigned the DSP filter forms with aligned columns, shorter "Low"/"High" labels, and Hz values right-aligned next to their units
  • Added buttons to add or remove an extra probe on a port from the Electrodes tab in Videre

Fixes

  • Fixed the live bandpass filter lagging one step behind when dragging the low-frequency slider
  • Fixed notch bandwidth rounding decimal values down (for example, 1.5 Hz became 1 Hz); it is now a Hz field with a unit
  • Fixed the bandpass "High" label selecting the "Low" field when clicked
  • Fixed fields being cut off in the DSP filter edit dialog
  • Removed the filter order shown for notch filters, which can't be changed
  • Fixed the Electrodes tab showing the first probe instead of the selected one when a port has several probes
  • Fixed overlapping labels and unstyled dropdowns in the Monitor view settings popover, and removed a detached arrow above it
  • Fixed the "Rows per page" selector in the Data Files table, and aligned and resized it to match its label
  • Fixed the "Data Files" toolbar's last button being pushed off the edge
  • Fixed spike sorter settings showing a value the server rejected instead of the value actually in use
  • Fixed the Spike Sorter tab repeatedly querying the server for a previous sort when no file was loaded
  • Fixed text in spike sorter gauges, raster and spike plots, the site activity heatmap, the neurons grid and PSTH showing in a serif font
  • Fixed the "Site Size" readout wrapping onto several lines in the Electrodes, Signal Metrics and 3D Model tabs
  • Changed the launcher's System icon label from "Signal Processing" to "System"
  • Changed the Analog Out Control table header from "Analog Out Channel" to "Channel"
  • Fixed the "+N" overflow count in multi-select pickers showing larger than the surrounding text

Allego 4.0.12

Released 2026-10-08 · Release on GitHub

Fixes

  • Fixed spike sorting ignoring the clustering settings (eps, min points, position scale) and using model defaults instead
  • Fixed a crash when switching simulated ports between channel counts (for example, 32 to 96 channels)
  • Fixed a crash when applying a notch filter with no bandwidth set; the setting is now rejected with an error
  • Fixed invalid spike width settings being reported as applied when the sorter rejected them
  • Fixed spike width changes having no effect on waveform-only sorting models
  • Fixed a stopped spike sort sometimes reporting as complete and saving its output
  • Fixed the spike sorter showing the wrong phase while waiting for enough spikes to train, and the dashboard not showing the sorter phase
  • Fixed raster plot data returning invalid timestamps when no end time was set
  • Fixed time-range selections that started partway through the recording returning data from the start
  • Fixed time-range selections made at a different sample rate returning the wrong samples

Allego 4.0.11

Released 2026-09-25 · Release on GitHub

Features

  • Renamed the "Invite User" button to "Invite Lab Member"
  • Renamed the invite dialog title from "Invite a User" to "Invite a Lab Member"
  • Added a help icon to the invite dialog that explains how lab seats and invitations work
  • Changed the invite confirmation message from "User invitation sent" to "Invitation sent"

Fixes

  • The "Invite Lab Member" button is now disabled when your license has no open seats or isn't active, instead of showing an error after you click it
  • Improved the styling of the invite form

Allego 4.0.2

Released 2026-07-01 · Release on GitHub

Features

  • Introduced feature-based licensing across the app — tabs, tools, and individual controls unlock according to your entitlements, with clear "locked" overlays where access is restricted.
  • Added an in-app auto-update system: a restart-to-update prompt, a manual Check for updates option, declinable updates, and reliable installation on macOS and Windows.
  • Added PSTH (peri-stimulus time histogram) analysis in the Neurons tab — multiple independent triggers, TTL and analog (AIN) trigger sources, configurable bin size and pre/post windows, per-trigger grouping, trial counts, pagination, and live auto-refresh.
  • Added a Spectral analysis tab with streaming/replay support, configurable frequency ranges, windowing options, an optional DSP-chain toggle, and image export.
  • Added loading of custom sensor/probe definitions from a file, with a guided modal and sensor-backup management (list, restore, delete).
  • Added a new a4x32_poly2_4mm probe.
  • Added a Signal Map button and modal to the Electrodes tab.
  • Added headstage global amp-settle stimulation controls.
  • Enhanced the HD Snapshot view with per-signal-type Y-axis ranges, keyboard navigation for time and amplitude, an independent step control, and a 4-mode channel-height selector.
  • Improved channel-group management with auto-generated names from channel ranges and enforced non-overlapping assignments.
  • Refreshed the look and feel with a new dark theme, updated typography (Inter and Fira Code), color-blind-friendly entity colors, and consistent theming that now carries into pop-out windows.
  • Reworked configuration into floating and slide-in overlay panels that minimize to chips, detach into full tabs, or dock back, with off-screen position clamping.
  • Added image download for analysis graphs and the 3D Model tab.
  • Added opt-in usage analytics with a privacy dialog so you control whether anonymous usage data is shared.
  • Enabled automatic fallback to simulator mode when hardware is unavailable at startup, instead of failing.

Fixes

  • Fixed multiple recording race conditions that could cause crashes, tail-data loss, or corrupted metadata, and improved recorder buffering for smoother real-time tailing.
  • Improved startup reliability so tabs populate their data even when the backend becomes available after the window loads.
  • Fixed stale headstage port-connection state after a headstage change.
  • Fixed DOUT channels incorrectly using DIN channel names for scaling.
  • Fixed spike raster data to return channel indices in channel mode and cluster labels in neuron mode.
  • Fixed "not found" responses surfacing as spurious error notifications.
  • Fixed macOS updates that downloaded successfully but failed to install.
  • Fixed stale signal-group selection after a datasource is invalidated.
  • Fixed stimulation channel range calculation for near-zero values and improved manual stim-trigger edge registration.
  • Fixed an "invalid append" error caused by duplicate data packets.
  • Resolved console errors and rendering glitches in the Electrodes, Spectral, and Neurons tabs, plus assorted plot/label issues.
  • Corrected date/time formatting to a consistent locale and improved file-duration and metadata display.
  • Improved reliability of loading saved workspaces so panel and tab state restore correctly.